21 CFR 820.181 : Food And Drugs (§ 1.1 To Parts 1403-1499) — (Subpart A To Parts 1272-1299) — Food And Drug Administration, Department Of Health And Human Services (Continued) (Subpart A To Parts 1272-1299) — Medical Devices (Subpart A To § 898.14) — Quality System Regulation (§ 820.1 To § 820.250) — Records (§ 820.180 To § 820.198) — Device master record.

4 Aug 2016

Each manufacturer shall maintain device master records (DMR's). Each manufacturer shall ensure that each DMR is prepared and approved in accordance with § 820.40. The DMR for each type of device shall include, or refer to the location of, the following information:

(a) Device specifications including appropriate drawings, composition, formulation, component specifications, and software specifications;

(b) Production process specifications including the appropriate equipment specifications, production methods, production procedures, and production environment specifications;

(c) Quality assurance procedures and specifications including acceptance criteria and the quality assurance equipment to be used;

(d) Packaging and labeling specifications, including methods and processes used; and

(e) Installation, maintenance, and servicing procedures and methods.

21 CFR 820.181