21 CFR 820.65 : Food And Drugs (§ 1.1 To Parts 1403-1499) — (Subpart A To Parts 1272-1299) — Food And Drug Administration, Department Of Health And Human Services (Continued) (Subpart A To Parts 1272-1299) — Medical Devices (Subpart A To § 898.14) — Quality System Regulation (§ 820.1 To § 820.250) — Identification And Traceability (§ 820.60 And § 820.65) — Traceability.

4 Aug 2016

Each manufacturer of a device that is intended for surgical implant into the body or to support or sustain life and whose failure to perform when properly used in accordance with instructions for use provided in the labeling can be reasonably expected to result in a significant injury to the user shall establish and maintain procedures for identifying with a control number each unit, lot, or batch of finished devices and where appropriate components. The procedures shall facilitate corrective action. Such identification shall be documented in the DHR.